Serum Metal Ion Levels Following Total Hip Arthroplasty With Modular Dual Mobility Components.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 26318084.
- Also identified by DOI 10.1016/j.arth.2015.07.035.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Dual mobility acetabular components can reduce the incidence of total hip arthroplasty (THA) instability. Modular dual mobility (MDM) components facilitate acetabular component implantation. However, corrosion can occur at modular junctions. Serum cobalt and chromium levels and Oxford scores were obtained at minimum two year follow-up for 100 consecutive patients who had THA with MDM components. Average Oxford score was 43 (range 13-48). Average serum cobalt and chromium values were 0.7 mcg/L (range, 0.0 to 7.0) and 0.6 mcg/L (range, 0.1 to 2.7), respectively. MARS MRI was performed for four patients with pain and elevated serum cobalt levels. Two of these studies were consistent with adverse local tissue reaction. We recommend use of MDM implants in only patients at high risk for dislocation following THA.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Chromium
- Cobalt
- Hip Prosthesis
Anatomy
- hip