Reducing contact forces in the arch and supra-aortic vessels using the Magellan robot.
other
Where this comes from
- Record sourced from PubMed, PMID 26386511.
- Also identified by DOI 10.1016/j.jvs.2015.06.215.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Conventional catheter manipulation in the arch and supra-aortic trunks carries a risk of cerebral embolization. This study proposes a platform for detailed quantitative analysis of contact forces (CF) exerted on the vasculature, in order to investigate the potential advantages of robotic navigation. An anthropomorphic phantom representing a type I bovine arch was mounted and coupled onto a force/torque sensor. Three-axis force readings provided an average root-mean-square modulus, indicating the total forces exerted on the phantom. Each of the left subclavian, left common carotid, and right common carotid arteries was cannulated within a simulated endovascular suite with conventional (n = 42) vs robotic techniques (n = 30) by two operator groups: experts and novices. The procedure path was divided into three phases, and performance metrics corresponding to mean and maximum forces, force impact over time, standard deviation of forces, and number of significant catheter contacts with the arterial wall were extracted. Overall, median CF were reduced from 1.20 N (interquartile range [IQR], 0.98-1.56 N) to 0.31 N (IQR, 0.26-0.40 N; P < .001) for the right common carotid artery; 1.59 N (IQR, 1.11-1.85 N) to 0.33 N (IQR, 0.29-0.43 N; P < .001) for the left common carotid artery; and 0.84 N (IQR, 0.47-1.08 N) to 0.10 N (IQR, 0.07-0.17 N; P < .001) for the left subclavian artery. Robotic navigation resulted in significant reductions for the mean and maximum forces for each procedural phase. Significant improvements were also seen in other metrics, particularly at the target vessel ostium and for the more anatomically challenging procedural phases. Force reductions using robotic technology were evident for both novice and expert groups. Robotic navigation can potentially reduce CF and catheter-tissue contact points in an in vitro model, by enhancing catheter stability and control during endovascular manipulation.
Medical subject headings
- Aorta, Thoracic
- Catheterization, Peripheral
- Embolism
- Endovascular Procedures
- Robotic Surgical Procedures
- Stroke
- Vascular Access Devices