Rectus-Sparing Technique for Driveline Insertion of Ventricular Assist Device.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 26522548.
- Also identified by DOI 10.1016/j.athoracsur.2015.02.138.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We present a new technique for driveline insertion of the HeartWare ventricular assist device (HVAD) designed to preserve the integrity of the abdominal wall structure. Because of the size of the HVAD driveline connector (12 mm in diameter: triple the size of the driveline cable), the standard tunneling maneuver can result in tearing of the abdominal wall muscle layer, which is a primary mechanism to prevent ascending driveline infection. We find that our technique is particularly useful in children because their abdominal wall muscles are more fragile and thereby prone to accidental injury with blunt penetration when the standard technique is used.
Medical subject headings
- Heart-Assist Devices
- Prosthesis Implantation
- Rectus Abdominis