A novel electromagnetic navigation tool for acetabular surgery.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 26542869.
- Also identified by DOI 10.1016/S0020-1383(15)30021-8.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Acetabular fracture surgery is demanding and screw placement along narrow bony corridors remains challenging. It necessitates x-ray radiation for fluoroscopically assisted screw insertion. The purpose of this cadaver study was to evaluate the feasibility, accuracy and operation time of a novel electromagnetic navigation system for screw insertion along predefined acetabular corridors. A controlled laboratory study with a total of 24 electromagnetically navigated screw insertions was performed on 8 cadaveric acetabula. 3 peri-acetabular bony corridors (QSS, Quadrilateral Surface Screw; IAS, Infra-Acetabular Screw; PCS, Posterior Column Screw) were defined and screws were placed in a defined order without fluoroscopy. Operation time was documented. Postoperative CT scans were performed to analyse accuracy of screw placement. Mean cadaver age was 70.4 ± 11.7. Successful screw placement was accomplished in 22 out of 24 (91.7%) cases. The overall mean time for all 3 acetabular screws was 576.6 ± 75.9s. All 3 complications occurred during the placement of the IAS due to an impassable narrow bony corridor. QSS mean length was 50 ± 5mm, IAS mean length was 85 ± 10mm and PCS mean length was 120 ± 5mm. In this cadaver study the novel electromagnetic navigation system was feasible to allow accurate screw placement without fluoroscopy in defined narrow peri-acetabular bony corridors.
Medical subject headings
- Acetabulum
- Fracture Fixation, Internal
- Fractures, Bone
- Surgery, Computer-Assisted
- Tomography, X-Ray Computed