Reliability of Pentacam HR Thickness Maps of the Entire Cornea in Normal, Post-Laser In Situ Keratomileusis, and Keratoconus Eyes.

Xu, Zhe; Peng, Mei; Jiang, Jun; Yang, Chun; Zhu, Weigen; Lu, Fan; Shen, Meixiao · Am J Ophthalmol · 2016

other · Level V

Where this comes from

Abstract

To measure the repeatability and reproducibility of Pentacam HR system thickness maps for the entire cornea in normal, post-laser in situ keratomileusis (post-LASIK), and keratoconus (KC) eyes. Reliability study. Sixty normal subjects (60 eyes), 30 post-LASIK subjects (60 eyes), and 14 KC patients (27 eyes) were imaged with the Pentacam HR system by 2 well-trained operators. For pachymetry the cornea was divided into 4 zones: a central zone (2-mm diameter) and concentric pericentral zone (2-5 mm), transitional zone (5-7 mm), and peripheral zone (7-10 mm). The 3 concentric zones were subdivided into 8 sectors. Intraobserver repeatability and interobserver reproducibility of entire corneal thickness maps were tested by the repeatability and reproducibility coefficients, intraclass correlation coefficients, coefficient of variation, and 95% limits of agreement. From central to peripheral zones, the precision of corneal thickness measurements became gradually smaller. Central zone repeatability and reproducibility were the best in the normal, post-LASIK, and KC groups. The peripheral superior sectors showed poorer repeatability and reproducibility for all subjects. The intraobserver repeatability and interobserver reproducibility for all zones were ≤19.3 μm, ≤22.1 μm, and ≤20.7 μm, in the normal, post-LASIK, and KC groups, respectively. The intraobserver and interobserver coefficients of variation for all zones were ≤1.3%, ≤1.6%, and ≤1.6% for all 3 groups. Pentacam HR system pachymetry of the entire cornea provided good precision in normal, post-LASIK, and KC corneas. Thickness measurements in the peripheral cornea should be interpreted with caution in abnormal corneas after surgery or with diseases.

Medical subject headings