Detection of volume loss using the Nexfin device in blood donors.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 26566706.
- Also identified by DOI 10.1111/anae.13283.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We investigated which haemodynamic parameters derived from Nexfin non-invasive continuous arterial blood pressure measurements are optimal to detect controlled volume loss in spontaneously breathing subjects. Haemodynamic monitoring was performed in 40 whole-blood donors. Mean arterial pressure, cardiac index, systemic vascular resistance index and pulse pressure variation were recorded during controlled breathing, and a Valsalva manoeuvre was performed before and after blood donation. Blood donation resulted in a reduction in cardiac index (from 3.96 ± 0.84 l.min(-1) .m(2) to 3.30 ± 0.61 l.min(-1) .m(2) ; p < 0.001), an increase in systemic vascular resistance (from 1811 ± 450 dyn.s.cm(-5) .m(2) to 2137 ± 428 dyn.s.cm(-5) .m(2) ; p < 0.001) and an increase in pulse pressure variation (from 13.4 ± 5.1 to 15.3 ± 5.4%; p = 0.02). The area under the receiver operating characteristic curve to detect volume loss was highest for cardiac index (0.94, 95% CI 0.88-0.99) and systemic vascular resistance (0.90, 95% CI 0.82-0.99). Nexfin is a non-invasive haemodynamic monitor that can feasibly detect volaemic changes in spontaneously breathing subjects.