Estimation of Fluorescence Lifetimes Via Rotational Invariance Techniques.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 26571506.
- Also identified by DOI 10.1109/TBME.2015.2491364.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Estimation of signal parameters via rotational invariance techniques is a classical algorithm widely used in array signal processing for direction-of-arrival estimation of emitters. Inspired by this method, a new signal model and new fluorescence lifetime estimation via rotational invariance techniques (FLERIT) were developed for multiexponential fluorescence lifetime imaging (FLIM) experiments. The FLERIT only requires a few time bins of a histogram generated by a time-correlated single-photon counting FLIM system, greatly reducing the data throughput from the imager to the signal processing units. As a noniterative method, the FLERIT does not require initial conditions, prior information nor model selection that are usually required by widely used traditional fitting methods, including nonlinear least square methods or maximum-likelihood methods. Moreover, its simplicity means it is suitable for implementations in embedded systems for real-time applications. FLERIT was tested on synthesized and experimental fluorescent cell data showing the potentials to be widely applied in FLIM data analysis.
Medical subject headings
- Image Processing, Computer-Assisted
- Microscopy, Fluorescence
- Signal Processing, Computer-Assisted