A pilot study of a hand-held camera in a busy burn centre: Prediction of patient length of recuperation with wound temperature.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 26654291.
- Also identified by DOI 10.1016/j.burns.2015.10.028.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of our study was to evaluate temperature differences of burns looking at their prognostic ability to predict healing at the 21 day mark. Thirty two burns in 26 patients aged 1-71 years old were photographed with a FLIR T650 camera. Environment, reflected, and body core temperature of the patients were measured. Skin emissivity was constant 0.98. Pictures were analyzed with R&D FLIR Software. Minimal and average burn temperatures and skin temperature in 255 pixel squares were measured. Patients were divided into healed and not healed groups. Statistical analysis was performed with SPSS 20 (IBM Armonk, USA) and p<0.05 was significant. There were 25 healed and 7 non-healed burns at 21 days. Healed burns were significantly warmer than non-healed burns (p<0.05). There was a statistically significant strong, negative correlation between the difference of minimal burns temperatures and healthy skin temperatures with days needed to heal the burns (p=0.001; rho=-0.564). Infrared camera seems to be useful equipment in predicting burns' healing time. However further clinical studies need to be done.
Medical subject headings
- Burns
- Infrared Rays
- Photography
- Skin Temperature
- Thermography
- Wound Healing