Thermal Damage Done to Bone by Burring and Sawing With and Without Irrigation in Knee Arthroplasty.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 26718777.
- Also identified by DOI 10.1016/j.arth.2015.11.002.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Heat from bone resecting tools used in knee surgery can induce thermal osteonecrosis, potentially causing aseptic implant loosening. This study compared oscillating saws to burrs in terms of temperature generation and histologic damage. Use of irrigation to reduce bone temperature was also investigated. Temperatures were recorded during sawing and burring with or without irrigation (uncooled or cooled). Histologic analyses were then carried out. Differences between groups were tested statistically (α = 0.05). On average, burring produced higher temperatures than sawing (P < .001). When uncooled irrigation was used, bone temperatures were significantly lower in sawed bone than in burred bone (P < .001). Irrigation lowered temperatures and thermal damage depths and increased osteocyte viability (P < .001). These results suggest that irrigating bone during resection could prevent osteonecrosis onset.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Bone and Bones
- Hot Temperature
- Osteonecrosis
- Therapeutic Irrigation
Anatomy
- knee