Stem Migration and Fretting Corrosion of the Antirotation Pin in the K2/Apex Hip System.

Kent, Michael; Edmondson, Mark; Ebert, Jay; Nivbrant, Nils; Kop, Alan; Wood, David; De Steiger, Richard · J Arthroplasty · 2016

rct · Level II

Where this comes from

Abstract

Many exchangeable neck hip systems have been withdrawn because of fretting corrosion at the neck/stem coupling. Our prospective randomized study evaluating stem stability (Roentgen stereophotogrammetric analysis, dual-energy x-ray absorptiometry) and clinical outcomes between the K2/Apex hip systems was ceased early because of a withdrawal of the stems which had an unfavorably high early revision rate reported in the Australian Orthopaedic Association National Joint Registry (9.3% at 3 years). At 2 years, there are no clinical differences between the stems. Roentgen stereophotogrammetric analysis has identified a high proportion of potentially concerning subsidence and retroversion in both groups, more marked in the K2 stem, although mostly in asymptomatic patients. Dual-energy x-ray absorptiometry has shown similar bone density around the stems. Retrieval analysis of 3 study patients showed fretting corrosion of the antirotation pin and aseptic lymphocyte-dominated vasculitis-associated lesion, with no relationship to bearing type or size. Analysis of 7 further nonstudy K2/Apex stems confirmed similar corrosion. This study shows potentially concerning subsidence of both stems and is the first to describe corrosion at the neck-stem interface and a relationship to metal-related pathology.