Prediction of patient-specific post-operative outcomes of TAVI procedure: The impact of the positioning strategy on valve performance.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 26748728.
- Also identified by DOI 10.1016/j.jbiomech.2015.10.048.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Prosthesis positioning in transcatheter aortic valve implantation procedures represents a crucial aspect for procedure success as demonstrated by many recent studies on this topic. Possible complications, device performance, and, consequently, also long-term durability are highly affected by the adopted prosthesis placement strategy. In the present work, we develop a computational finite element model able to predict device-specific and patient-specific replacement procedure outcomes, which may help medical operators to plan and choose the optimal implantation strategy. We focus in particular on the effects of prosthesis implantation depth and release angle. We start from a real clinical case undergoing Corevalve self-expanding device implantation. Our study confirms the crucial role of positioning in determining valve anchoring, replacement failure due to intra or para-valvular regurgitation, and post-operative device deformation.
Medical subject headings
- Patient-Specific Modeling
- Transcatheter Aortic Valve Replacement