Feasibility of Dose Optimization in a Second-Generation Dual-Source CT Scanner for a Manufacturer-Recommended Urolithiasis Protocol for Imaging Renal Stones.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 26797362.
- Also identified by DOI 10.2214/AJR.15.15204.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this article is to investigate the magnitude of dose optimization for a manufacturer-recommended urolithiasis protocol in a second-generation dual-source CT scanner. Custom renal phantoms with 24 stones were scanned using the manufacturer-provided dual-energy CT protocol (tube A, 100 kVp and 210 reference mAs; tube B, 140 kVp and 162 reference mAs) and seven dose-optimized protocols in which the reference tube current-time product setting of tube A was reduced stepwise by 20 mAs. Detection and characterization of the stones was assessed. In the patient study, 25 patients underwent the manufacturer-provided dual-energy protocol and 25 patients underwent imaging with a dose-optimized protocol (tube A, 100 kVp and 90 reference mAs; tube B, 140 kVp and 70 reference mAs). Dose-length product (DLP), image noise, and contrast-to-noise ratio (CNR) were assessed. Subjective image quality was analyzed by three independent radiologists. In the phantom study, the reference tube current-time product of tube A could be reduced from 210 to 90 mAs without losing the accuracy of detection or characterization of the calculi. In the patient study, the dose-optimized protocol resulted in a significant reduction of the average DLP by 51% compared with the standard protocol (219.4 vs 443.5 mGy·cm, respectively; p = 0.0001). The image noise was higher, and the CNR was lower, in the dose-optimized group than in the standard-dose group (p < 0.05). The subjective overall image quality of the dose-optimized CT examinations was rated as good, and that of the standard-dose CT examinations was rated as excellent (p = 0.001). The in vitro and in vivo assessment revealed a potential for a 51% dose reduction of the manufacturer-recommended dual-energy CT protocol for urolithiasis without compromising the accuracy.
Medical subject headings
- Kidney Calculi
- Tomography, X-Ray Computed