Wear study of Total Ankle Replacement explants by microstructural analysis.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 26807869.
- Also identified by DOI 10.1016/j.jmbbm.2015.12.035.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The implantation of Total Ankle Replacement (TAR) prostheses generally gives satisfactory results. However, a high revision rate is associated with the Ankle Evolutive System (AES) implant, due to periprosthetic osteolysis that generates significant cortical lesions and bone cysts in the periprosthetic region. Radioclinical and histological analyses of peri-implant tissues show the presence of numerous foreign particles that may come from the implant. It is known that a precocious wear of materials may lead to an important rate of foreign body in tissues and may generate osteolysis lesions and inflammatory reactions. Thus the objectives of this retrospective study of 10 AES TAR implants (recovered after revision surgeries) are to understand how the prostheses wear out, which part is the most stressed and to determine the nature and size of foreign body particles. A better understanding of friction mechanisms between the three parts of the implant and of the nature and morphology of foreign particles generated was needed to explain the in vivo behavior of the implant. This was achieved using microstuctural and tomographic analysis of both implants parts and periprosthetic tissues.
Medical subject headings
- Arthroplasty, Replacement, Ankle
- Inflammation
- Joint Prosthesis
- Osteolysis
- Prosthesis Failure