To evaluate, in the FLEX Registry, clinical outcomes of an ultrathin (60 µm) biodegradable polymer-coated Supraflex sirolimus-eluting stent (SES) for the treatment of coronary artery disease. Additionally, to determine the vascular response to the Supraflex SES through optical coherence tomography (OCT) analysis. Multicentre, single-arm, all-comers, observational registry of patients who were treated with the Supraflex SES, between July 2013 and May 2014, at nine different centres in India. 995 patients (1242 lesions) who were treated with the Supraflex SES, between July 2013 and May 2014, at nine different centres in India. A total of 47 participants underwent OCT analysis at 6 months' follow-up. Percutaneous coronary intervention with Supraflex SES, PRIMARY AND SECONDARY OUTCOME MEASURES: The primary endpoint-the rate of major adverse cardiac events (defined as a composite of cardiac death, myocardial infarction (MI), target lesion revascularisation (TLR))-was analysed during 12 months. At 12 months, the primary endpoint occurred in 36 (3.7%) of 980 patients, consisting of 18 (1.8%) cardiac deaths, 16 (1.6%) MI, 7 (0.7%) TLR and 2 (0.2%) cases of non-target lesion target vessel revascularization. In a subset of 47 patients, 1227 cross-sections (9309 struts) were analysed at 6 months by OCT. Overall, a high percentage of struts was covered (98.1%), with a mean neointimal thickness of 0.13 ± 0.06 µm. The FLEX Registry evaluated clinical outcomes in real-world and more complex cohorts and thus provides evidence that the Supraflex SEX can be used safely and routinely in a broader percutaneous coronary intervention population. Also, the Supraflex SES showed high percentage of stent strut coverage and good stent apposition during OCT follow-up.
Medical subject headings
- Antibiotics, Antineoplastic
- Coronary Artery Disease
- Drug-Eluting Stents
- Percutaneous Coronary Intervention
- Registries
- Sirolimus