Squeaking and microcracks in a delta-delta ceramic coupling: pin-on-disc study.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 26971327.
- Also identified by DOI 10.1007/s00590-016-1748-0.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
There is a rising concern about squeaking in ceramic-on-ceramic total hip arthroplasty (THA). In pin-on-disc testing of a delta-delta coupling, we reproduced squeaking and observed microcracks on worn surfaces. We used a pin-on-disc machine and made discs and pins by cutting delta ceramic to a diameter of 40 mm (D-D). Cross-linked polyethylene was used for a comparison disc (D-P). We performed the same test using another D-D coupling specimen to confirm reproducibility. Squeaking in the D-D specimen was reproduced in wet conditions, though it was not found in the D-P specimen. Fast Fourier transform analysis showed a peak frequency for squeaking of 2794 Hz. The noise occurred at about 6.6 km of sliding distance. Scanning electron microscopy revealed that the worn surface of the pin in D-D at 10.8 km of sliding distance had some microcracks. However, there was no obvious damage to the worn surface of the pin in D-P at the same sliding distance. We confirmed the reproducibility of these findings, obtaining similar results, including squeaking, from another D-D coupling specimen. Our findings show that squeaking may occur in THA using delta ceramic bearings even if implants are placed to avoid extra-articular impingement of the femoral neck. Although the clinical relevance of microcracks is unknown, they may affect long-term outcomes in THA using delta ceramic bearings.
Medical subject headings
- Hip Prosthesis
Anatomy
- hip