The effectiveness of pulse oximetry sonification enhanced with tremolo and brightness for distinguishing clinically important oxygen saturation ranges: a laboratory study.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 26990198.
- Also identified by DOI 10.1111/anae.13424.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Our study examined the effectiveness of pulse oximetry sonification enhanced with acoustic tremolo and brightness to help listeners differentiate clinically relevant oxygen saturation ranges. In a series of trials lasting 30 s each, 76 undergraduate participants identified final oxygen saturation range ( 100% to 97%; Low: 96% to 90%; Critical: 89% and below), and detected threshold transitions into and out of the target range using conventional sonification (n = 38) or enhanced sonification (n = 38). Median (IQR [range]) accuracy for range identification with the conventional sonification was 80 (70-85 [45-95])%, whereas with the enhanced sonification it was 100 (99-100 [80-100])%; p < 0.001. Accuracy for detecting threshold transitions with the conventional sonification was 60 (50-75 [30-95])%, but with the enhanced sonification it was 100 (95-100 [75-100]%; p < 0.001. Participants can identify clinically meaningful oxygen saturation ranges and detect threshold transitions more accurately with enhanced sonification than with conventional sonification.
Medical subject headings
- Oximetry
- Oxygen