Once Annealed Highly Cross-Linked Polyethylene Exhibits Low Wear at 9 to 15 Years.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 27088350.
- Also identified by DOI 10.3928/01477447-20160414-03.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A once annealed highly cross-linked polyethylene (HXLPE) was introduced in 1998. Concerns regarding its long-term performance and oxidative resistance exist because of the presence of retained free radicals. The authors studied 48 patients with 50 hip implants having an average age of 62 years. They were followed for 9 to 15 years. The purpose of this study was to determine linear wear rate and the incidence of osteolysis and/or mechanical failure. At an average follow-up of 12.2 years, the annual linear wear rate was 0.018 mm (SD, 0.024 mm). No mechanical failures or osteolysis have been found to date. The clinical performance of this HXLPE continues to meet expectations despite the presence of free radicals. [Orthopedics. 2016; 39(3):e565-e571.].
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
- Polyethylene
- Prosthesis Failure