Use of three-dimensional digital image correlation to evaluate mechanical response of prosthetic systems.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 27091866.
- Also identified by DOI 10.1177/0309364616637956.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The need of comfortable and safe prosthetic systems is an important challenge for both prosthetists and engineers. The aim of this technical note is to demonstrate the use of three-dimensional digital image correlation to evaluate mechanical response of two prosthetic systems under real patient dynamic loads. This note describes the use of three-dimensional digital image correlation method to obtain full-field strain and displacement measurements on the surface of two lower limb prostheses for Chopart amputation. It outlines key points of the measurement protocol and illustrates the analysis of critical regions using data obtained on specific points of interest. The results show that the use of three-dimensional digital image correlation can be a tool for the prosthetist to optimize the prosthesis considering features related to the material and design, in order to bear with real patient-specific load conditions. Clinical relevance Three-dimensional digital image correlation can support decision-making on new designs and materials for prosthetics based on quantitative data. Better understanding of mechanical response could also assist prescription for appropriate prosthetic systems.
Medical subject headings
- Artificial Limbs
- Imaging, Three-Dimensional
- Prosthesis Design