Highly Sensitive Detection of Melamine Using a One-Step Sample Treatment Combined with a Portable Ag Nanostructure Array SERS Sensor.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 27120183.
- Also identified by DOI 10.1371/journal.pone.0154402 and PMC identifier 4847794.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
There is an urgent need for rapid and reliable methods able to detect melamine in animal feed. In this study, a quick, simple, and sensitive method for the determination of melamine content in animal feed was developed using surface-enhanced Raman spectroscopy on fabricated Ag nanorod (AgNR) array substrates with a one-step sample extraction procedure. The AgNR array substrates washed by HNO3 solvent (10-7 M) and methanol and showed the good stability within 6 months. The Raman shift at △ν = 682 cm-1 was used as the characteristic melamine peak in the calculations. Sufficient linearity was obtained in the 2-200 μg·g-1 range (R2 = 0.926). The limits of detection and quantification were 0.9 and 2 μg·g-1, respectively. The recovery rates were 89.7-93.3%, with coefficients of variation below 2.02%. The method showed good accuracy compared with the tradition GC-MS analysis. This new protocol only need 2 min to fininsh the detection which could be developed for rapid onsite screening of melamine contamination in quality control and market surveillance applications.
Medical subject headings
- Animal Feed
- Chemistry Techniques, Analytical
- Metal Nanoparticles
- Resins, Synthetic
- Triazines