Failure mechanisms in CoCrMo modular femoral stems for revision total hip arthroplasty.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 27122233.
- Also identified by DOI 10.1002/jbm.b.33693.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In this retrieval study, we reported the failure mechanisms of the CoCrMo-based hip implants. Systematic analyses on the clinically failed modular femoral stems from Revitan™ revision prostheses revealed a multistep fracture process. Multiple microcracks were first developed under the combined action of pitting corrosion and dynamic tensile stress on the lateral side of the CoCrMo connection taper. These microcracks then served as the initiation sites of further corrosion fatigue cracking leading to the final catastrophic failure. This crack initiation process has not been previously reported on retrieved CoCrMo components and our findings provide valuable information on the clinical performance of such implants, as well as the material selection and structural designs of future modular stems. © 2016 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 105B: 1525-1535, 2017.
Medical subject headings
- Alloys
- Arthroplasty, Replacement, Hip
- Chromium
- Cobalt
- Equipment Failure Analysis
- Hip Prosthesis
- Molybdenum
- Prosthesis Failure