Integration of Immature Granulocytes Quantification With the Version 2.0 UniCel DxH 800 in the HematoFlow Strategy.
Where this comes from
- Record sourced from PubMed, PMID 27124946.
- Also identified by DOI 10.1093/ajcp/aqw035.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Our aim was to define whether the early granulocyte cell marker (EGC%_DxH) parameter might replace immature granulocytes counts obtained by HematoFlow (IG%_HF) and/or manual differential count (IG%_manual). We conducted a study over a 10-day period in February 2014 whereby 402 samples were analyzed for the IG flag. We correlated the EGC%_DxH vs IG%_HF and IG%_manual, identified any discrepant results and finally looked at the impact on our workflow by incorporation of the EGC% into our WBC differential algorithm. On an initial training set, a receiver operating characteristic (ROC) curve analysis showed a threshold of 0.9% for EGC%_DxH (sensitivity of 91.7%, specificity of 93.5% and an area under the curve of 0.965). Further analysis of the dataset (259 samples) found a correlation of the EGC%_DxH to all our IG% counting methods (r = 0.963). Incorporation of the EGC%_DxH into the WBC HematoFlow differential resulted in a 36% reduction of samples requiring HematoFlow and/or slide review. The EGC% generated by the DxH 800 can be easily incorporated into existing HematoFlow and slide review algorithms.
Medical subject headings
- Algorithms
- Biomarkers
- Granulocytes
- Immunophenotyping
- Leukocyte Count