A theoretical model to predict tensile deformation behavior of balloon catheter.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 27214691.
- Also identified by DOI 10.1016/j.jmbbm.2016.04.044.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In this technical note, a simple theoretical model was proposed to express the tensile deformation and fracture of balloon catheter tested by the ISO standard using piece-wise linear force-displacement relations. The model was then validated by comparing with the tensile force-displacement behaviors of two types of typical balloon catheters clinically used worldwide. It was shown that the proposed model can effectively be used to express the tensile deformation behavior and easily be handled by physicians who are not familiar with mechanics of materials.
Medical subject headings
- Catheters
- Models, Theoretical
- Stress, Mechanical