A Novel Device for the Evaluation of Hemostatic Function in Critical Care Settings.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 27224934.
- Also identified by PMC identifier 5536171.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Major surgical procedures often result in significant intra- and postoperative bleeding. The ability to identify the cause of the bleeding has the potential to reduce the transfusion of blood products and improve patient care. We present a novel device, the Quantra Hemostasis Analyzer, which has been designed for automated, rapid, near-patient monitoring of hemostasis. The Quantra is based on Sonic Estimation of Elasticity via Resonance Sonorheometry, a proprietary technology that uses ultrasound to measure clot time and clot stiffness from changes in viscoelastic properties of whole blood during coagulation. We present results of internal validation and analytical performance testing of the technology and demonstrate the ability to characterize the key functional components of hemostasis.
Medical subject headings
- Blood Coagulation Tests
- Critical Care
- Hemostasis
- Rheology
- Ultrasonics