Perforation forces of the intact porcine anterior lens capsule.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 27254279.
- Also identified by DOI 10.1016/j.jmbbm.2016.05.007.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
During the first step of cataract surgery, the lens capsule is perforated and a circular hole is created with a sharp instrument, a procedure called capsulorhexis. To develop automated systems that can assist ophthalmologists during capsulorhexis, the forces required must be quantified. This study investigates perforation forces of the central anterior lens capsule in porcine eyes, which are used as a conservative model for the human eye. A micro-mechanical characterisation method is presented that measures capsular bag perforation forces with a high precision positioning and high-resolution force sensing system. The force during perforation of the anterior lens capsule was measured with various sized needles and indentation speeds and is found to be 15-35mN. A bio-mechanical model is identified that describes an exponential correlation between indentation force and depth, indicating strain hardening behaviour of the porcine anterior lens capsule.
Medical subject headings
- Capsulorhexis
- Lens Capsule, Crystalline