Feasibility of self-gated isotropic radial late-phase MR imaging of the liver.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 27271925.
- Also identified by DOI 10.1007/s00330-016-4433-0.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate feasibility of a 3D-isotropic self-gated radial volumetric interpolated breath-hold examination (VIBE) for late-phase MRI of the liver. 70 patients were included and underwent liver MRI at 1.5 T. Depending on the diagnosis, either Gd-EOB-DTPA (35 patients) or gadobutrol (35 patients) were administered. During late (gadobutrol) or hepatocyte-specific phase (Gd-EOB-DTPA), a radial prototype sequence was acquired and reconstructed using (1) self-gating with 40 % acceptance (rVIBE<sub>40</sub>); (2) with 100 % acceptance of the data (rVIBE<sub>100</sub>) and compared to Cartesian VIBE (cVIBE). Images were assessed qualitatively (image quality, lesion conspicuity, artefacts; 5-point Likert-scale: 5 = excellent; two independent readers) and quantitatively (coefficient-of-variation (CV); contrast-ratio) in axial and coronal reformations. In eight cases only rVIBE provided diagnostic image quality. Image quality of rVIBE<sub>40</sub> was rated significantly superior (p < 0.05) in Gd-EOB-DTPA-enhanced and coronal reformatted examinations as compared to cVIBE. Lesion conspicuity was significantly improved (p < 0.05) in coronal reformatted Gd-EOB-DTPA-enhanced rVIBE<sub>40</sub> in comparison to cVIBE. CV was higher in rVIBE<sub>40</sub> as compared to rVIBE<sub>100</sub>/cVIBE (p < 0.01). Gadobutrol-enhanced rVIBE<sub>40</sub> and cVIBE showed higher contrast-ratios than rVIBE<sub>100</sub> (p < 0.001), whereas no differences were found in Gd-EOB-DTPA-enhanced examinations. Self-gated 3D-isotropic rVIBE provides significantly superior image quality compared to cVIBE, especially in multiplanar reformatted and Gd-EOB-DTPA-enhanced examinations. • Radial VIBE acquisition reduces motion artefacts. • Gd-EOB-DTPA-enhanced scans provide improved image quality. • Non-diagnostic liver MRI examinations may be reduced by radial k-spaces sampling.
Medical subject headings
- Image Interpretation, Computer-Assisted
- Imaging, Three-Dimensional
- Liver Diseases
- Magnetic Resonance Imaging