Meeting The Joint Commission's Dose Incident Identification and External Benchmarking Requirements Using the ACR's Dose Index Registry.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 27318581.
- Also identified by DOI 10.1016/j.jacr.2016.04.026.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this investigation was to evaluate the potential of using the ACR's Dose Index Registry(®) to meet The Joint Commission's requirements to identify incidents in which the radiation dose index from diagnostic CT examinations exceeded the protocol's expected dose index range. In total, 10,970 records in the Dose Index Registry were statistically analyzed to establish both an upper and lower expected dose index for each protocol. All 2015 studies to date were then retrospectively reviewed to identify examinations whose total examination dose index exceeded the protocol's defined upper threshold. Each dose incident was then logged and reviewed per the new Joint Commission requirements. Facilities may leverage their participation in the ACR's Dose Index Registry to fully meet The Joint Commission's dose incident identification review and external benchmarking requirements.
Medical subject headings
- Benchmarking
- Joint Commission on Accreditation of Healthcare Organizations
- Radiation Monitoring
- Registries
- Risk Management
- Tomography, X-Ray Computed