Assessment of failure of cemented polyethylene acetabular component due to bone remodeling: A finite element study.

Ghosh, Rajesh · J Orthop · 2016

biomechanical · Level V

Where this comes from

Abstract

The aim of the study is to determine failure of the cemented polyethylene acetabular component, which might occur due to excessive bone resorption, cement-bone interface debonding and fatigue failure of the cement mantle. Three-dimensional finite element models of intact and implanted pelvic bone were developed and bone remodeling algorithm was implemented for present analysis. Soderberg fatigue failure diagram was used for fatigue assessment of the cement mantle. Hoffman failure criterion was considered for prediction of cement-bone interface debonding. Results indicate fatigue failure of the cement mantle and implant-bone interface debonding might not occur due to bone remodeling.

Anatomy