Efficacy Assessment of Nucleic Acid Decontamination Reagents Used in Molecular Diagnostic Laboratories.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 27410228.
- Also identified by DOI 10.1371/journal.pone.0159274 and PMC identifier 4943653.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
The occurrence of nucleic acid cross contamination in the laboratory resulting in false positive results of diagnostic samples is seriously problematic. Despite precautions to minimize or even avoid nucleic acid cross contaminations, it may appear anyway. Until now, no standardized strategy is available to evaluate the efficacy of commercially offered decontamination reagents. Therefore, a protocol for the reliable determination of nucleic acid decontamination efficacy using highly standardized solution and surface tests was established and validated. All tested sodium hypochlorite-based reagents proved to be highly efficient in nucleic acid decontamination even after short reaction times. For DNA Away, a sodium hydroxide-based decontamination product, dose- and time-dependent effectiveness was ascertained. For two other commercial decontamination reagents, the phosphoric acid-based DNA Remover and the non-enzymatic reagent DNA-ExitusPlus™ IF, no reduction of amplifiable DNA/RNA was observed. In conclusion, a simple test procedure for evaluation of the elimination efficacy of decontamination reagents against amplifiable nucleic acid is presented.
Medical subject headings
- Decontamination
- Equipment Contamination
- Pathology, Molecular
- Sodium Hydroxide
- Sodium Hypochlorite