Total arthroplasty of basal thumb joint with Elektra prothesis: an in vitro analysis.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 27424207.
- Also identified by DOI 10.1177/1753193416659230.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The reported outcomes of the Elektra thumb carpo-metacarpal joint implant have been very variable. This study evaluates the influence of daily cyclic loads and the type of the screw-fit cup insertion technique in the trapezium, with and without prior threading, on the structural bone behaviour. The study was performed experimentally to predict initial implant stability and cortical bone strains. Computational models were developed to assess the structural cancellous bone behaviour. The use of Elektra implant considerably changed the bone strain behaviour compared with the intact joint. This may be associated with risks of cancellous bone fatigue failure due to overload, particularly in the trapezium. The joint load magnitude has a more important structural role than that of the screw-fit cup insertion technique. Limiting the magnitude of thumb loads after arthroplasty may contribute positively to the longevity of this procedure. V.
Medical subject headings
- Arthroplasty, Replacement
- Carpometacarpal Joints
- Joint Prosthesis
- Osteoarthritis
- Thumb
- Trapezium Bone