Quantitative MRI assessment of a novel direction modulated brachytherapy tandem applicator for cervical cancer at 1.5T.

Soliman, Abraam S; Elzibak, Alyaa; Easton, Harry; Kim, James Y; Han, Dae Yup; Safigholi, Habib; Mashouf, Shahram; Owrangi, Amir et al. · Radiother Oncol · 2016

biomechanical · Level V

Where this comes from

Abstract

The purpose of this work is to quantitatively investigate the artifacts and image distortions induced in the MR images by a recently proposed direction modulated brachytherapy (DMBT) tandem applicator prototype. This new MRI-compatible applicator allows better sparing of organs-at-risk (OAR) for cervical cancer patients, while providing conformal dose distributions to target volumes. Specific phantom and tools were designed and manufactured for this study. The phantom was filled with a tissue-like solution and MR images were acquired with clinical protocols as per GEC-ESTRO recommendations. Images were obtained at 6 different orientations that mimic possible clinical settings and full-width-at-half-maximum (FWHM) was recorded at multiple locations/angles. The accuracy of detecting the centerline of the tandem was assessed using a novel radial-fiducials mount. FWHM from all line profiles at all angles and all orientations was 6.14±0.7mm (compared to 6mm of the actual DMBT tandem diameter). The in-plane spatial-shift observed at para-axial and para-sagittal views was less than 0.5mm. This work demonstrated that the novel DMBT tandem applicator prototype has minimal artifact in T<sub>2</sub>-weighted images employed in clinical practice, suggesting the applicator might be a good candidate for MRI-guided adaptive brachytherapy.