Liquid fiducial marker performance during radiotherapy of locally advanced non small cell lung cancer.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 27443450.
- Also identified by DOI 10.1016/j.radonc.2016.06.012.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We analysed the positional and structural stability of a long-term biodegradable liquid fiducial marker (BioXmark) for radiotherapy in patients with locally advanced lung cancer. Markers were injected via endoscopic- or endobronchial ultrasound in lymph nodes and reachable primary tumours. Marker volume and Hounsfield Units (HU) changing rates were estimated using breath-hold CBCT. Inter-fraction variation in marker position relative to gross tumour volume (GTV) position was established, as well as the inter-fraction variation in mediastinal marker registration relative to a carina registration through the treatment. Fifteen patients were included and 29 markers analysed. All markers that were in situ at planning were visible through the treatment. Mean HU was 902±165HU for lymph node and 991±219HU for tumour markers. Volume degradation rates were -5% in lymph nodes and -23% in primary tumours. Three-dimensional inter-fraction variation for marker position relative to the GTV position was -0.1±0.7mm in lymph nodes and -1.5±2.3mm in primary tumours. Inter-fraction variations in marker registration relative to carina registration were -0.4±1.2mm in left-right, 0.2±2.0mm in anterior-posterior and -0.5±2.0mm in cranio-caudal directions. The liquid fiducial markers were visible and stable in size and position throughout the treatment course.
Medical subject headings
- Carcinoma, Non-Small-Cell Lung
- Fiducial Markers
- Lung Neoplasms
- Radiotherapy Planning, Computer-Assisted