Berlin Heart EXCOR Ventricular Assist Device: Multilayer Membrane Rupture in a Pediatric Patient.

Di Molfetta, Arianna; Filippelli, Sergio; Ferrari, Gianfranco; Secinaro, Aurelio; Zielinski, Krystzof; Amodeo, Antonio · Ann Thorac Surg · 2016

case_report · Level V

Where this comes from

Abstract

A 2-year-old child was implanted with an Berlin Heart EXCOR Ventricular Assist Device (Berlin Heart, Berlin, Germany) as a bridge to heart transplantation for idiopathic dilated cardiomyopathy. At postoperative day 296, a significant reduction of membrane movement was observed. The device was explanted and tested on a hydronumerical circulation simulator. Findings suggested that the integrity of the multilayered membrane had been compromised. This was confirmed by a computed tomography scan of the device. The computed tomography evidenced a detachment of the 3-layered membrane, with a thinner, convex layer on the side of the air chamber and an opposite convexity of the remaining membranes. These showed an additional air space within the layers.