Effects of Dual-Energy Technique on Radiation Exposure and Image Quality in Pediatric Body CT.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 27490819.
- Also identified by DOI 10.2214/AJR.15.15994.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to assess the effects of dual-energy CT (DECT) on radiation exposure and image quality in pediatric body CT. This retrospective study included 79 children (median age, 10.1 years; range, 12 days-18 years) who underwent thoracic or abdominal-pelvic CT or CT angiography with dual-energy technique between October 2014 and March 2015. The delivered volume CT dose index (CTDI<sub>vol</sub>) from DECT was recorded and compared with the estimated CTDI<sub>vol</sub> had the patient undergone scanning with a standard single-energy CT (SECT) protocol. Size-specific dose estimates were calculated for both DECT and SECT. Image quality was subjectively scored (scale, 1-4). For 16 of 79 patients who underwent both DECT and SECT, image contrast and noise were measured and contrast-to-noise ratio calculated. Parametric and nonparametric testing of independent and paired samples was performed. For all 79 studies, actual median CTDI<sub>vol</sub> and size-specific dose estimate were 3.7 and 5.9 mGy for DECT versus prescanning estimates of 4.4 and 7.7 mGy for SECT, resulting in 12.5% and 11.2% radiation exposure reduction (p < 0.01). Diagnostic image quality was achieved in all patients. In the 16-patient subset, the median CTDI<sub>vol</sub> values of DECT and SECT were 3.1 and 3.4 mGy (p < 0.05). Median noise was greater with DECT than with SECT (p < 0.01), but the mean contrast-to-noise ratios for the liver and portal vein were similar (liver, p = 0.32; portal vein, p = 0.21). In pediatric body CT, the use of DECT results in radiation exposures comparable to or less than those of SECT while maintaining contrast and contrast-to-noise ratio.