Digital palpation of the pilot balloon vs. continuous manometry for controlling the intracuff pressure in laryngeal mask airways.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 27501056.
- Also identified by DOI 10.1111/anae.13566.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study compared two methods of controlling the intracuff pressure in laryngeal mask airways. One hundred and eighty patients were randomly assigned into two groups. In the first group (n = 90), after training, the intracuff pressure was controlled using digital palpation of the pilot balloon. In the second group (n = 90), continuous manometry was used to control the intracuff pressure. An upper pressure limit of 60 cmH2 O was set. The median (IQR [range]) intracuff pressure in the palpation group was 130 (125-130 [120-130]) cmH2 O compared with 29 (20-39 [5-60]) cmH2 O in the manometry group (p < 0.001). In the palpation group, 37% of patients experienced pharyngolaryngeal complications vs. 12% in the manometry group (p < 0.001). We conclude that the digital palpation technique is not a suitable alternative to manometry in controlling the intracuff pressure in laryngeal mask airways.
Medical subject headings
- Laryngeal Masks
- Monitoring, Physiologic
- Palpation
- Postoperative Complications