Cyclic Fatigue Resistance and Force Generated by OneShape Instruments during Curved Canal Preparation.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 27513666.
- Also identified by DOI 10.1371/journal.pone.0160815 and PMC identifier 4981425.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
OBJECTIVES: To evaluate the cyclic fatigue resistance and the force generated by OneShape files during preparation of simulated curved canals. METHODS: Six OneShape files (the test) and six ProTaper F2 files (the control) were subject to the bending ability test. Another thirty files of each type were used to prepare artificial canals (n = 60), which were divided into 3 groups according to respective curvatures of the canals (30°, 60°, and 90°). The numbers of cycles to fatigue (NCF) as well as the positive and negative forces that were generated by files during canal preparation were recorded. The scanning electron microscopy was applied to detect the fracture surfaces. RESULTS: Compared with ProTaper F2 files, the bending loads of OneShape files were significantly lower at deflections of 45°(P < .05), 60° (P < .05) and 75° (P < .01). No significant difference was found at 30°. OneShape files presented a higher NCF in both 60° and 90° canals than the control (P < .01). No significant difference of NCF was found between OneShape and ProTaper files in 30° canals. During the preparation of 30° canals by both files, the negative forces were dominant. With the increase of the curvature, more positive forces were observed. When the OneShape Files were compared with the control, significant different forces were found at D3 and D2 (P < .05) in 30° canals, at D2 (P < .05), D1 (P < .01) and D0 (P < .01) in 60° canals, and at D4 and D3 (P < .01) in 90° canals. CONCLUSIONS: OneShape files possessed a reliable flexibility and cyclic fatigue resistance. According to the assessments of the forces generated by files, OneShape instruments performed in a more fatigue-resistant way during curved canal preparation, compared with the ProTaper F2 files.
Medical subject headings
- Dental Instruments
- Equipment Design
- Materials Testing
- Root Canal Preparation