Transcatheter Valve-in-Valve: A Cautionary Tale.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 27549545.
- Also identified by DOI 10.1016/j.athoracsur.2016.01.045.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Transcatheter aortic valve replacement (TAVR) by valve-in-valve (VIV) implantation is an alternative treatment for high-risk patients with a degenerating aortic bioprosthesis. We present a case of transapical TAVR VIV with a 29-mm Edwards SAPIEN XT (ESV) (Edwards Lifesciences, Irvine, CA) into a 29-mm Medtronic Freestyle stentless bioprosthesis (Medtronic Inc, Minneapolis, MN) in which unanticipated dilatation of the Freestyle bioprosthesis resulted in intraprocedural embolization of the TAVR valve, necessitating urgent conversion to a conventional surgical aortic valve replacement (AVR). Our experience suggests that TAVR VIV with the 29-mm ESV in the setting of a degenerated 29-mm Freestyle stentless bioprosthesis must be undertaken with caution.
Medical subject headings
- Aortic Valve Stenosis
- Bioprosthesis
- Coronary Stenosis
- Intraoperative Complications
- Thoracotomy
- Transcatheter Aortic Valve Replacement