Relationship between pull-out strength and oscillation angle in bipolar cups: an in vitro study.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 27574265.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To compare the past and present bipolar hip arthroplasty (BHA) models in terms of balance between pull-out strength and oscillation angle (OA). The pull-out strength and OA of 8 BHA models were compared: UPF-II, IBC, and Tandem XLPE (Smith & Nephew); Ringloc x (Biomet); J-FX (DePuy); Bipolar (Nakashima Medical); Multipolar (Zimmer); and Centrax (Stryker). Respectively for the UPF-II, IBC, Tandem, Ringloc, J-FX, Nakashima Bipolar, Multipolar, and Centrax, the mean pull-out strength was 2219 N, 3303 N, 1503 N, 951 N, 1453 N, 1856 N, 1536 N, and 753 N, whereas the mean OA was 54.2°, 53.8°, 64.0°, 73.2°, 63.0°, 65.4°, 55.6°, and 75.4°. The OA was lower in the integrated types. For pull-out strength of the locking mechanism, the integrated type (IBC and Nakashima) was stronger than the metal or polyethylene ring-lock type (all others). The pull-out strength and OA were negatively correlated (r= -0.881, p=0.007), and the balance between the 2 varied for different models. There is a trade-off between the pullout strength and OA; optimal balance between the 2 should be based on each patient's need.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis