Safety of LigaSure in recurrent laryngeal nerve dissection-porcine model using continuous monitoring.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 27578605.
- Also identified by DOI 10.1002/lary.26271.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study investigated recurrent laryngeal nerve (RLN) real-time electromyography (EMG) data to define optimal safety parameters of the LigaSure Small Jaw (LSJ) instrument during thyroidectomy. Prospective animal model. Dynamic EMG tracings were recorded from 32 RLNs (16 piglets) during various applications of LSJ around using continuous electrophysiologic monitoring. At varying distances from the RLN, the LSJ was activated (activation study). The LSJ was also applied to the RLN at timed intervals after activation and after a cooling maneuver through placement on the sternocleidomastoid muscle (cooling study). In the activation study, there was no adverse EMG event at 2 to 5 mm distance (16 RLNs, 96 tests). In the cooling study, there was no adverse EMG event after 2-second cooling time (16 RLNs, 96 tests) or after the LSJ cooling maneuver on the surrounding muscle before reaching the RLNs (8 RLNs, 24 tests). Based on EMG functional assessment, the safe distance for LSJ activation was 2 mm. Further LSJ-RLN contact was safe if the LSJ was cooled for more than 2 seconds or cooled by touch muscle maneuver. The LSJ should be used with these distance and time parameters in mind to avoid RLN injury. N/A. Laryngoscope, 127:1724-1729, 2017.
Medical subject headings
- Dissection
- Electromyography
- Monitoring, Intraoperative
- Recurrent Laryngeal Nerve
- Surgical Instruments
- Thyroidectomy