Radiation exposure using the O-arm<sup>®</sup> surgical imaging system.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 27652675.
- Also identified by DOI 10.1007/s00586-016-4773-0.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study was conducted to characterise the O-arm<sup>®</sup> surgical imaging system in terms of patient organ doses and medical staff occupational exposure during three-dimensional thoracic spine and pelvic examinations. An anthropomorphic phantom was used to evaluate absorbed organ doses during a three-dimensional thoracic spine scan and a three-dimensional pelvic scan with the O-arm<sup>®</sup>. Staff occupational exposure was evaluated by constructing an ambient dose cartography of the operating theatre during a three-dimensional pelvic scan as well as using an anthropomorphic phantom to simulate the O-arm<sup>®</sup> operator. Patient organ doses ranged from 30 ± 4 μGy to 20.0 ± 3.0 mGy and 4 ± 1 μGy to 6.7 ± 1.0 mGy for a three-dimensional thoracic spine and pelvic examination, respectively. For a single three-dimensional acquisition, the maximum ambient equivalent dose at 2 m from the iso-centre was 11 ± 1 μSv. Doses delivered to the patient during a three-dimensional thoracic spine image acquisition were found to be significant with the O-arm<sup>®</sup>, but lower than those observed with a standard computed tomography examination. The detailed dose cartography allows for the optimisation of medical staff positioning within the operating theatre while imaging with the O-arm<sup>®</sup>.
Medical subject headings
- Fluoroscopy
- Occupational Exposure
- Radiation Dosage
- Radiation Exposure