Late Outcomes of Transcatheter Aortic Valve Replacement in High-Risk Patients: The FRANCE-2 Registry.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 27712776.
- Also identified by DOI 10.1016/j.jacc.2016.07.747.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Transcatheter aortic valve replacement (TAVR) has revolutionized management of high-risk patients with severe aortic stenosis. However, survival and the incidence of severe complications have been assessed in relatively small populations and/or with limited follow-up. This report details late clinical outcome and its determinants in the FRANCE-2 (FRench Aortic National CoreValve and Edwards) registry. The FRANCE-2 registry prospectively included all TAVRs performed in France. Follow-up was scheduled at 30 days, at 6 months, and annually from 1 to 5 years. Standardized VARC (Valve Academic Research Consortium) outcome definitions were used. A total of 4,201 patients were enrolled between January 2010 and January 2012 in 34 centers. Approaches were transarterial (transfemoral 73%, transapical 18%, subclavian 6%, and transaortic or transcarotid 3%) or, in 18% of patients, transapical. Median follow-up was 3.8 years. Vital status was available for 97.2% of patients at 3 years. The 3-year all-cause mortality was 42.0% and cardiovascular mortality was 17.5%. In a multivariate model, predictors of 3-year all-cause mortality were male sex (p < 0.001), low body mass index, (p < 0.001), atrial fibrillation (p < 0.001), dialysis (p < 0.001), New York Heart Association functional class III or IV (p < 0.001), higher logistic EuroSCORE (p < 0.001), transapical or subclavian approach (p < 0.001 for both vs. transfemoral approach), need for permanent pacemaker implantation (p = 0.02), and post-implant periprosthetic aortic regurgitation grade ≥2 of 4 (p < 0.001). Severe events according to VARC criteria occurred mainly during the first month and subsequently in <2% of patients/year. Mean gradient, valve area, and residual aortic regurgitation were stable during follow-up. The FRANCE-2 registry represents the largest database available on late results of TAVR. Late mortality is largely related to noncardiac causes. Incidence rates of severe events are low after the first month. Valve performance remains stable over time.
Medical subject headings
- Aortic Valve Stenosis
- Transcatheter Aortic Valve Replacement