Coiling of wide-necked carotid artery aneurysms assisted by a temporary bridging device (Comaneci): preliminary experience.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 27742747.
- Also identified by DOI 10.1136/neurintsurg-2016-012664 and PMC identifier 5740552.
- Licence recorded as CC BY-NC.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Endovascular treatment of wide-necked aneurysms remains challenging without the use of adjunctive devices to preserve the parent artery. To present our initial experience with a temporary bridging device, the Comaneci (Rapid Medical, Israel). The compliant remodeling mesh protects the parent artery during coil occlusion without flow arrest in the distal vasculature. Permanent dual antiplatelet therapy is not required since the device is fully removed at the end of the procedure. All intracranial aneurysms treated by coil occlusion with the Comaneci device between December 2014 and November 2015 were included. Angiographic and clinical results were retrospectively analysed, including follow-up examinations. All aneurysms were unruptured. 18 aneurysms of the internal carotid artery were included. Successful coil occlusion assisted by the Comaneci device as intended was possible in 14 cases (77.8%). Insufficient coverage of the aneurysmal neck was observed in four cases (22.2%), with modification of the treatment to stent-assisted coiling or remodeling. One clinically relevant complication occurred (5.6%). Our initial experience shows that use of the Comaneci device is straightforward for the treatment of selected wide-necked aneurysms. Further studies with long-term follow-up data are needed to identify the significance of the presented technique in the neurointerventional armamentarium.
Medical subject headings
- Carotid Artery, Internal
- Intracranial Aneurysm
- Surgical Mesh