Contamination of single-use bronchoscopes in critically ill patients.
Where this comes from
- Record sourced from PubMed, PMID 27747865.
- Also identified by DOI 10.1111/anae.13622.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Disposable bronchoscopes such as the Ambu aScope<sup>TM</sup> 3 are marketed as 'single use' The risks of contamination from prolonged device storage before possible re-use are unknown. Following clinical bronchoscopy in patients whose lungs were mechanically ventilated, 20 aScope<sup>TM</sup> 3's bronchoscopes received a standard 'social clean' and were then stored. Subsequent paired saline flush and swab samples were taken at time zero, and at 24 h and 48 h. Positive microbiological cultures were obtained from at least one time point from 16 of the 20 bronchoscopes. Pathogens considered at high risk of causing pneumonia were isolated from seven bronchoscopes, with significant quantities from six of them. Our study demonstrates that aScope<sup>TM</sup> 3's should not be re-used on the same patient, as clinically significant growth of micro-organisms occurs frequently, despite adequate social cleaning. Culture of bronchoscopes themselves may be a potentially useful diagnostic tool in the context of pulmonary infection. Our data make it clear that these devices are single use and not single patient use.
Medical subject headings
- Bronchoscopes
- Critical Illness
- Cross Infection
- Disposable Equipment
- Equipment Contamination