Head-neck taper corrosion following total hip arthroplasty with Stryker Meridian stem.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 27768219.
- Also identified by DOI 10.5301/hipint.5000440.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Taper corrosion at the modular junctions of total hip arthroplasty (THA) femoral stems are known to cause locally destructive adverse local tissue reaction (ALTR). However, the implants at risk remain to be fully elucidated. We report the case of a 76-year-old woman with hip pain, abductor weakness, and sciatic nerve symptoms 2 years following metal-on-polyethylene THA. Serum cobalt levels were elevated, while chromium levels were normal. Magnetic resonance imaging demonstrated ALTR. The patient's symptoms resolved following head and liner exchange with ceramic head and titanium sleeve. Previous studies have reported ALTR secondary to head-neck taper corrosion in cobalt chrome alloy and titanium alloy stems used in metal-on-polyethylene (MoP) THA. The stems described above span 4 manufacturers with varying taper geometries. To our knowledge, this is the first report of ALTR due to head-neck taper corrosion in a Stryker Meridian titanium-molybdenum-zirconium-iron (TMZF) alloy stem. Close surveillance for ALTR may be considered in patients with this prosthesis. Further investigation of the TMZF alloy and V40 taper geometry of this stem may guide future implant design.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
- Joint Diseases
- Postoperative Complications
- Prosthesis Failure
Anatomy
- hip