Comparative performance assessment of commercially available automatic external defibrillators: A simulation and real-life measurement study of hands-off time.

Savastano, Simone; Vanni, Vincenzo; Burkart, Roman; Raimondi, Maurizio; Canevari, Fabrizio; Molinari, Simone; Baldi, Enrico; Danza, Aurora I et al. · Resuscitation · 2017

cross_sectional · Level IV

Where this comes from

Abstract

Early and good quality cardiopulmonary resuscitation (CPR) and the use of automated external defibrillators (AEDs) improve cardiac arrest patients' survival. However, AED peri- and post-shock/analysis pauses may reduce CPR effectiveness. The time performance of 12 different commercially available AEDs was tested in a manikin based scenario; then the AEDs recordings from the same tested models following the clinical use both in Pavia and Ticino were analyzed to evaluate the post-shock and post-analysis time. None of the AEDs was able to complete the analysis and to charge the capacitors in less than 10s and the mean post-shock pause was 6.7±2.4s. For non-shockable rhythms, the mean analysis time was 10.3±2s and the mean post-analysis time was 6.2±2.2s. We analyzed 154 AED records [104 by Emergency Medical Service (EMS) rescuers; 50 by lay rescuers]. EMS rescuers were faster in resuming CPR than lay rescuers [5.3s (95%CI 5-5.7) vs 8.6s (95%CI 7.3-10). AEDs showed different performances that may reduce CPR quality mostly for those rescuers following AED instructions. Both technological improvements and better lay rescuers training might be needed.