Comparative performance assessment of commercially available automatic external defibrillators: A simulation and real-life measurement study of hands-off time.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 27780740.
- Also identified by DOI 10.1016/j.resuscitation.2016.10.006.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Early and good quality cardiopulmonary resuscitation (CPR) and the use of automated external defibrillators (AEDs) improve cardiac arrest patients' survival. However, AED peri- and post-shock/analysis pauses may reduce CPR effectiveness. The time performance of 12 different commercially available AEDs was tested in a manikin based scenario; then the AEDs recordings from the same tested models following the clinical use both in Pavia and Ticino were analyzed to evaluate the post-shock and post-analysis time. None of the AEDs was able to complete the analysis and to charge the capacitors in less than 10s and the mean post-shock pause was 6.7±2.4s. For non-shockable rhythms, the mean analysis time was 10.3±2s and the mean post-analysis time was 6.2±2.2s. We analyzed 154 AED records [104 by Emergency Medical Service (EMS) rescuers; 50 by lay rescuers]. EMS rescuers were faster in resuming CPR than lay rescuers [5.3s (95%CI 5-5.7) vs 8.6s (95%CI 7.3-10). AEDs showed different performances that may reduce CPR quality mostly for those rescuers following AED instructions. Both technological improvements and better lay rescuers training might be needed.
Medical subject headings
- Cardiopulmonary Resuscitation
- Defibrillators
- Electric Countershock
- Emergency Medical Services
- First Aid
- Out-of-Hospital Cardiac Arrest