Metal ion levels in ceramic-on-ceramic THR with cobalt-chrome modular necks: analysis of cobalt and chromium serum levels in 23 healthy hip patients.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 27791242.
- Also identified by DOI 10.5301/hipint.5000430.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Modular necks in total hip replacement (THR) can be a source of metal ion release. We measured the level of cobalt (Co) and chromium (Cr) ions in the serum of 23 healthy volunteers with a well-functioning ceramic-on-ceramic THR of a particular design, containing a cobalt-chrome (Co-Cr) stem and a Co-Cr modular neck. Average Co in serum was 1,71 µg/l; median Co was 1.50 µg/l. No patient had Co levels higher than 3.70 µg/l. Average Cr level was 0.49 µg/l; in all but one patient Cr was below the detection limit (<0.50 µg/l). Co was higher for varus necks compared to all other subtypes: 2.14 µg/l vs 1.58 µg/l (p<0.05). The same system with Titanium modular necks shows lower Co serum levels: 1.26 µg/l (p<0.01 ). In conclusion, the hybrid THR Profemur Xm® - Procotyle L® with a Co-Cr modular neck on a Co-Cr stem design shows very low Cr ion serum levels in a randomly selected group of well-functioning hip patients. Co ion serum levels are low, but significantly higher for 8° varus necks, and Co levels are significantly higher than values of this system with a Titanium modular neck.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Ceramics
- Osteoarthritis, Hip
- Prosthesis Failure
Anatomy
- hip