National Registry Data and Record Linkage to Inform Postmarket Surveillance of Prosthetic Aortic Valve Models Over 15 Years.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 27820610.
- Also identified by DOI 10.1001/jamainternmed.2016.6936.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Postmarket evidence generation for medical devices is important yet limited for prosthetic aortic valve devices in the United Kingdom. To identify prosthetic aortic valve models that display unexpected patterns of mortality or reintervention using routinely collected national registry data and record linkage. This observational study used data from all National Health Service and private hospitals in England and Wales that submit data to the National Adult Cardiac Surgery Audit (NACSA). All patients undergoing first-time elective and urgent aortic valve replacement surgery (with or without coronary artery bypass grafting) with a biological (n = 15 series) or mechanical (n = 10 series) prosthetic valve from 5 primary suppliers, and satisfying prespecified data quality criteria (n = 43 782 biological; n = 11 084 mechanical) between 1998 and 2013 were included. Valves were classified into series of related models. Outcome tracking was performed using multifaceted record linkage. The median follow-up was 4.1 years (maximum, 15.3 years). Cox proportional hazards regression with random effects (frailty models) were used to model valve effects on the outcomes, with and without adjustment for preoperative and intraoperative covariates. Time to all-cause mortality or aortic valve reintervention (surgical or transcatheter). There were 13 104 deaths and 723 reinterventions during follow-up. Of 79 345 isolated aortic valve replacement procedures with or without coronary artery bypass grafting, 54 866 were analyzed. Biological valve implantation rates increased from 59% in 1998 and 1999 to 86% in 2012 and 2013. Two series of valves associated with significantly increased hazard of death or reintervention were identified (first series: frailty, 1.18; 95% prediction interval [PI], 1.06-1.32 and second series: frailty, 1.19; 95% PI, 1.09-1.31). These results were robust to covariate adjustment and sensitivity analyses. There were 3 prosthetic valves with a significant reduction in hazard (valve 1: frailty, 0.88; 95% PI, 0.80-0.96; valve 2: frailty, 0.88; 95% PI, 0.80-0.96; and valve 3: frailty, 0.88; 95% PI, 0.78-0.98). Meaningful evidence from the analysis of routinely collected registry data can inform postmarket surveillance of medical devices. Although the findings are associated with a number of caveats, 2 specific biological aortic valve series identified in this study may warrant further investigation.
Medical subject headings
- Aortic Valve
- Heart Valve Prosthesis
- Heart Valve Prosthesis Implantation
- Medical Record Linkage
- Product Surveillance, Postmarketing
- Registries