A Patient-Specific Instrument for Femoral Stem Placement During Total Hip Arthroplasty.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 27841929.
- Also identified by DOI 10.3928/01477447-20161108-06.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To ensure that the femoral stem is placed in the proper position during total hip arthroplasty, the authors developed a patient-specific instrument. A total of 10 total hip arthroplasties were performed with the assistance of the patient-specific instrument during this study. The mean accuracy of stem tilt, varus/valgus, and anteversion was 2.1°±4.1°, 1.0°±0.7°, and 4.7°±1.2°, respectively. No complications were observed and no reoperations were required for any of the patients who underwent surgery included in this study. The results support the feasibility of this patient-specific instrument for use during stem placement in total hip arthroplasty. [Orthopedics. 2017; 40(2):e374-e377.].