Valvular Regurgitation After Implantation of Prostheses Secured With Cor-Knot Automated Fasteners.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 27899293.
- Also identified by DOI 10.1016/j.athoracsur.2016.11.035.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Valvular regurgitation is a recognized complication of aortic and mitral valve procedures, with paravalvular leak the primary mechanism and transvalvular leak less common. The Cor-Knot automated fastener (LSI Solutions, Victor, NY) has been shown to be a safe, viable alternative to knot pushers or manual tying for use in these procedures. Here we present two patients in whom the orientation of the Cor-Knot fastener appeared to result in transvalvular regurgitation necessitating a redo operation. With adjunctive use of Cor-Knot devices, fastener orientation relative to valve geometry must be assessed for optimal results.
Medical subject headings
- Aortic Valve Insufficiency
- Heart Valve Prosthesis Implantation
- Mitral Valve Insufficiency
- Postoperative Complications
- Surgical Fixation Devices