10-year follow-up after implantation of the Bryan Cervical Disc Prosthesis.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 27904963.
- Also identified by DOI 10.1007/s00586-016-4897-2.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Cervical arthroplasty is being used as an alternative for cervical fusion, but long-term follow-up results have rarely been reported. In this paper, we present 10-year follow-up results after implantation of the Bryan Cervical Disc Prosthesis in a single center. 89 patients underwent implantation of a single-level Bryan Cervical Disc Prosthesis to treat radiculopathy and/or myelopathy. Clinical (Neurological Success, Neck Disability Index (NDI), Neck- and Arm-Pain, and SF-36) and radiological follow-up was prospectively organized up to 10 years after surgery. Adverse events and second surgeries were recorded and evaluated. Ten-year follow-up data were available for 72 (81%) patients. Maintenance or improvement of the neurological state was seen in 89% of patients after 10-year follow-up. SF-36 PCS scores improved significantly at all follow-up points. SF-36 MCS improvement was significant at 4 and 6 year, but not at 8- and 10-year follow-up. Significant improvement for NDI, and Neck- and Arm-Pain scores was found for the subgroup of patients in whom these data were available. Mean angular motion of the prosthesis at 10-year follow-up was 8.6°. Mobility of the device, defined as >2° of angular motion, was reached in 81% of patients. During the study period, 21 patients (24%) developed new or recurrent radiculopathy or myelopathy, the majority of these being treated conservatively. Seven patients (8%) required 8 additional spine surgeries to treat persistent or recurrent symptoms. Of these, 2 patients (2%) were reoperated at the index level and at 5 (6%) an adjacent level. In this study, favorable long-term clinical outcome after implantation of the Bryan Cervical Disc Prosthesis was seen, with the majority of prostheses remaining mobile after 10-year follow-up. However, still 6% of patients required adjacent level surgery.
Medical subject headings
- Cervical Vertebrae
- Intervertebral Disc
- Orthopedic Procedures
- Prosthesis Implantation
Anatomy
- cervical spine