Metaphyseal locking plate as an external fixator for open tibial fracture: Clinical outcomes and biomechanical assessment.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 27919511.
- Also identified by DOI 10.1016/j.injury.2016.11.031.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study aimed to evaluate the outcome of using a metaphyseal locking plate as a definitive external fixator for treating open tibial fractures based on biomechanical experiments and analysis of clinical results. A metaphyseal locking plate was used as an external fixator in 54 open tibial fractures in 52 patients. The mean follow-up was 38 months (range, 20-52 months). Moreover, static axial compression and torsional tests were performed to evaluate the strength of the fixation techniques. The average fracture healing time was 34.5 weeks (range, 12-78 weeks). At 4 weeks postoperatively and at the final follow-up, the average Hospital for Special Surgery knee score was 85 (range, 81-100) and 94 (range, 88-100), respectively, and the American Orthopaedic Foot and Ankle Society score was 88 (range, 80-100) and 96 (range, 90-100), respectively. Based on the static test result, the axial stiffness was significantly different among groups (p=0.002), whereas the torsional stiffness showed no significant difference (p=0.068). Clinical outcomes show that the use of locking plate as a definitive external fixator is an alternative choice for tibial fractures after obtaining appropriate fracture reduction. However, external locked plating constructs were not as strong as standard locked plating constructs. Therefore, the use of external locked plating constructs as a definitive treatment warrants further biomechanical study for construct strength improvement.
Medical subject headings
- Fracture Fixation
- Fractures, Open
- Radiography
- Tibial Fractures